Search
Find a vulnerability
Search criteria
7 vulnerabilities
CVE-2026-97361 (GCVE-0-2026-97361)
Vulnerability from cvelistv5 – Published: 2026-11-13 17:57 – Updated: 2026-11-16 07:47
VLAI
EPSS
VEX
Title
Improper Privilege Management on B. Braun SpaceCom, Battery Pack SP with Wi-Fi
Summary
There exists an improper privilege management vulnerability that could cause privilege escalation at the SpaceCom and Space Battery pack SP with WiFi.
Severity
SSVC
Exploitation: none
Automatable: no
Technical Impact: total
CISA Coordinator · CISA-ADP (v2.0.3)
Decision recorded 2026-11-13 18:29 UTC
CWE
- CWE-269 - Improper Privilege Management
Assigner
References
2 references
Impacted products
2 products
| Vendor | Product | Version | |
|---|---|---|---|
| B. Braun Melsungen AG | SpaceCom |
Affected:
0 , ≤ 011L000094
(custom)
Affected: 0 , ≤ 012U000094 (custom) |
|
| B. Braun Melsungen AG | Space Battery Pack SP with Wi-Fi |
Affected:
0 , ≤ 053L000094
(custom)
Affected: 0 , ≤ 054U000094 (custom) |
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2026-97361",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2026-11-13T18:29:56.225817Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2026-11-13T19:31:35.380Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"product": "SpaceCom",
"vendor": "B. Braun Melsungen AG",
"versions": [
{
"lessThanOrEqual": "011L000094",
"status": "affected",
"version": "0",
"versionType": "custom"
},
{
"lessThanOrEqual": "012U000094",
"status": "affected",
"version": "0",
"versionType": "custom"
}
]
},
{
"defaultStatus": "unaffected",
"product": "Space Battery Pack SP with Wi-Fi",
"vendor": "B. Braun Melsungen AG",
"versions": [
{
"lessThanOrEqual": "053L000094",
"status": "affected",
"version": "0",
"versionType": "custom"
},
{
"lessThanOrEqual": "054U000094",
"status": "affected",
"version": "0",
"versionType": "custom"
}
]
}
],
"credits": [
{
"lang": "en",
"type": "finder",
"value": "Codewhite GmbH, Germany"
}
],
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cspan\u003eThere exists an improper privilege management vulnerability that could cause privilege escalation at the \u003c/span\u003e\u003cspan\u003eSpaceCom\u003c/span\u003e\u003cspan\u003e and \u003c/span\u003e\u003cspan\u003eSpace Battery pack \u003c/span\u003e\u003cspan\u003eSP \u003c/span\u003e\u003cspan\u003ewith \u003c/span\u003e\u003cspan\u003eWiFi\u003c/span\u003e\u003cspan\u003e.\u003c/span\u003e"
}
],
"value": "There exists an improper privilege management vulnerability that could cause privilege escalation at the SpaceCom and Space Battery pack SP with WiFi."
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "NONE",
"attackVector": "LOCAL",
"baseScore": 8.5,
"baseSeverity": "HIGH",
"exploitMaturity": "NOT_DEFINED",
"privilegesRequired": "LOW",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "NONE",
"subConfidentialityImpact": "NONE",
"subIntegrityImpact": "NONE",
"userInteraction": "NONE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:L/AC:L/AT:N/PR:L/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N",
"version": "4.0",
"vulnAvailabilityImpact": "HIGH",
"vulnConfidentialityImpact": "HIGH",
"vulnIntegrityImpact": "HIGH",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-269",
"description": "CWE-269 Improper Privilege Management",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2026-11-16T07:47:07.639Z",
"orgId": "653264ec-f98b-4e8f-b8b4-540a01b7657d",
"shortName": "B.Braun"
},
"references": [
{
"url": "https://www.bbraun.com/productsecurity"
},
{
"url": "https://www.bbraunusa.com/productsecurity"
}
],
"solutions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "The following versions include a patch for this vulnerability:\u003cdiv\u003e\u003cul\u003e\u003cli\u003eSpaceCom: 011L000096 or later / 012U000096 or later,\u003c/li\u003e\u003cli\u003e\n\u003cspan\u003eBattery Pack SP with WiFi: 053L000095 or later / 054U000095 or later\u003c/span\u003e\n\n\u003c/li\u003e\u003c/ul\u003e\u003c/div\u003e"
}
],
"value": "The following versions include a patch for this vulnerability: * SpaceCom: 011L000096 or later / 012U000096 or later,\n * \nBattery Pack SP with WiFi: 053L000095 or later / 054U000095 or later"
}
],
"source": {
"discovery": "EXTERNAL"
},
"title": "Improper Privilege Management on B. Braun SpaceCom, Battery Pack SP with Wi-Fi",
"x_generator": {
"engine": "Vulnogram 0.2.0"
}
}
},
"cveMetadata": {
"assignerOrgId": "653264ec-f98b-4e8f-b8b4-540a01b7657d",
"assignerShortName": "B.Braun",
"cveId": "CVE-2026-97361",
"datePublished": "2026-11-13T17:57:12.207Z",
"dateReserved": "2026-09-24T13:33:19.163Z",
"dateUpdated": "2026-11-16T07:47:07.639Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.2"
}
CVE-2026-97358 (GCVE-0-2026-97358)
Vulnerability from cvelistv5 – Published: 2026-11-13 17:55 – Updated: 2026-11-16 07:46
VLAI
EPSS
VEX
Title
Inclusion of Web Functionality from an Untrusted Source on B. Braun SpaceCom, Battery Pack SP with Wi-Fi
Summary
An inclusion of web functionality from an untrusted source could allow remote code execution on SpaceCom
Severity
SSVC
Exploitation: none
Automatable: no
Technical Impact: total
CISA Coordinator · CISA-ADP (v2.0.3)
Decision recorded 2026-11-13 18:31 UTC
CWE
- CWE-830 - Inclusion of Web Functionality from an Untrusted Source
Assigner
References
2 references
Impacted products
1 product
| Vendor | Product | Version | |
|---|---|---|---|
| B. Braun Melsungen AG | SpaceCom |
Affected:
0 , ≤ 011L000094
(custom)
Affected: 0 , ≤ 012U000094 (custom) |
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2026-97358",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2026-11-13T18:31:26.488418Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2026-11-13T19:31:51.522Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"product": "SpaceCom",
"vendor": "B. Braun Melsungen AG",
"versions": [
{
"lessThanOrEqual": "011L000094",
"status": "affected",
"version": "0",
"versionType": "custom"
},
{
"lessThanOrEqual": "012U000094",
"status": "affected",
"version": "0",
"versionType": "custom"
}
]
}
],
"credits": [
{
"lang": "en",
"type": "finder",
"value": "Codewhite GmbH, Germany"
}
],
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cspan\u003eAn inclusion of web functionality from an untrusted source \u003c/span\u003e\u003cspan\u003ecould \u003c/span\u003e\u003cspan\u003eallow remote code execution \u003c/span\u003e\u003cspan\u003eon\u003c/span\u003e\u003cspan\u003e \u003c/span\u003e\u003cspan\u003eSpaceCom\u003c/span\u003e"
}
],
"value": "An inclusion of web functionality from an untrusted source could allow remote code execution on SpaceCom"
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "PRESENT",
"attackVector": "NETWORK",
"baseScore": 7,
"baseSeverity": "HIGH",
"exploitMaturity": "NOT_DEFINED",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "LOW",
"subConfidentialityImpact": "LOW",
"subIntegrityImpact": "HIGH",
"userInteraction": "NONE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:L/VI:L/VA:L/SC:L/SI:H/SA:L",
"version": "4.0",
"vulnAvailabilityImpact": "LOW",
"vulnConfidentialityImpact": "LOW",
"vulnIntegrityImpact": "LOW",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-830",
"description": "CWE-830\tInclusion of Web Functionality from an Untrusted Source",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2026-11-16T07:46:12.542Z",
"orgId": "653264ec-f98b-4e8f-b8b4-540a01b7657d",
"shortName": "B.Braun"
},
"references": [
{
"url": "https://www.bbraun.com/productsecurity"
},
{
"url": "https://www.bbraunusa.com/productsecurity"
}
],
"solutions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cspan\u003eThe following versions include a patch for this vulnerability:\u0026nbsp;\n\u003cspan\u003eSpaceCom: 011L000096 or later / 012U000096.\u003c/span\u003e\u003c/span\u003e"
}
],
"value": "The following versions include a patch for this vulnerability:\u00a0\nSpaceCom: 011L000096 or later / 012U000096."
}
],
"source": {
"discovery": "UNKNOWN"
},
"title": "Inclusion of Web Functionality from an Untrusted Source on B. Braun SpaceCom, Battery Pack SP with Wi-Fi",
"x_generator": {
"engine": "Vulnogram 0.2.0"
}
}
},
"cveMetadata": {
"assignerOrgId": "653264ec-f98b-4e8f-b8b4-540a01b7657d",
"assignerShortName": "B.Braun",
"cveId": "CVE-2026-97358",
"datePublished": "2026-11-13T17:55:24.556Z",
"dateReserved": "2026-09-24T13:21:42.792Z",
"dateUpdated": "2026-11-16T07:46:12.542Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.2"
}
CVE-2026-97357 (GCVE-0-2026-97357)
Vulnerability from cvelistv5 – Published: 2026-11-13 17:54 – Updated: 2026-11-16 07:47
VLAI
EPSS
VEX
Title
Out-of-bounds write on B. Braun SpaceCom, Battery Pack SP with Wi-Fi
Summary
An Out-Of-Bounds Write due to missing input validation is possible in the autoprogramming interface of SpaceCom and Space Battery pack with WiFi which could lead to a denial of service of the affected communication device
Severity
SSVC
Exploitation: none
Automatable: yes
Technical Impact: partial
CISA Coordinator · CISA-ADP (v2.0.3)
Decision recorded 2026-11-13 18:31 UTC
CWE
- CWE-787 - Out-of-bounds Write
Assigner
References
2 references
Impacted products
2 products
| Vendor | Product | Version | |
|---|---|---|---|
| B. Braun Melsungen AG | SpaceCom |
Affected:
0 , ≤ 011L000094
(custom)
Affected: 0 , ≤ 012U000094 (custom) |
|
| B. Braun Melsungen AG | Space Batterypack SP with Wi-Fi |
Affected:
0 , ≤ 053L000094
(custom)
Affected: 0 , ≤ 054U000094 (custom) |
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2026-97357",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "yes"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2026-11-13T18:31:52.934925Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2026-11-13T19:32:09.161Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"product": "SpaceCom",
"vendor": "B. Braun Melsungen AG",
"versions": [
{
"lessThanOrEqual": "011L000094",
"status": "affected",
"version": "0",
"versionType": "custom"
},
{
"lessThanOrEqual": "012U000094",
"status": "affected",
"version": "0",
"versionType": "custom"
}
]
},
{
"defaultStatus": "unaffected",
"product": "Space Batterypack SP with Wi-Fi",
"vendor": "B. Braun Melsungen AG",
"versions": [
{
"lessThanOrEqual": "053L000094",
"status": "affected",
"version": "0",
"versionType": "custom"
},
{
"lessThanOrEqual": "054U000094",
"status": "affected",
"version": "0",
"versionType": "custom"
}
]
}
],
"credits": [
{
"lang": "en",
"type": "finder",
"value": "Codewhite GmbH, Germany"
}
],
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cspan\u003eAn Out-Of-Bounds Write due to missing input validation is possible in the autoprogramming interface of SpaceCom and Space Battery pack with WiFi which could lead to a denial of service of the affected communication device\u003c/span\u003e\n\n\u003cspan\u003e\u003c/span\u003e"
}
],
"value": "An Out-Of-Bounds Write due to missing input validation is possible in the autoprogramming interface of SpaceCom and Space Battery pack with WiFi which could lead to a denial of service of the affected communication device"
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "NONE",
"attackVector": "NETWORK",
"baseScore": 6.9,
"baseSeverity": "MEDIUM",
"exploitMaturity": "NOT_DEFINED",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "NONE",
"subConfidentialityImpact": "NONE",
"subIntegrityImpact": "NONE",
"userInteraction": "NONE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:N/VI:L/VA:L/SC:N/SI:N/SA:N",
"version": "4.0",
"vulnAvailabilityImpact": "LOW",
"vulnConfidentialityImpact": "NONE",
"vulnIntegrityImpact": "LOW",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-787",
"description": "CWE-787 Out-of-bounds Write",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2026-11-16T07:47:48.690Z",
"orgId": "653264ec-f98b-4e8f-b8b4-540a01b7657d",
"shortName": "B.Braun"
},
"references": [
{
"url": "https://www.bbraun.com/productsecurity"
},
{
"url": "https://www.bbraunusa.com/productsecurity"
}
],
"solutions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cdiv\u003eThe following versions contain a patch to this vulnerability:\u0026nbsp;\u003c/div\u003e\u003cdiv\u003e\u003cul\u003e\u003cli\u003e\u003cspan\u003eSpaceCom: 011L000096 or later / 012U000096 or later,\u003c/span\u003e\u003c/li\u003e\u003cli\u003eBattery Pack SP with WiFi: 053L000095 / 054U000095 or later.\u003c/li\u003e\u003c/ul\u003e\u003c/div\u003e"
}
],
"value": "The following versions contain a patch to this vulnerability:\u00a0\n\n * SpaceCom: 011L000096 or later / 012U000096 or later,\n * Battery Pack SP with WiFi: 053L000095 / 054U000095 or later."
}
],
"source": {
"discovery": "UNKNOWN"
},
"title": "Out-of-bounds write on B. Braun SpaceCom, Battery Pack SP with Wi-Fi",
"x_generator": {
"engine": "Vulnogram 0.2.0"
}
}
},
"cveMetadata": {
"assignerOrgId": "653264ec-f98b-4e8f-b8b4-540a01b7657d",
"assignerShortName": "B.Braun",
"cveId": "CVE-2026-97357",
"datePublished": "2026-11-13T17:54:23.896Z",
"dateReserved": "2026-09-24T13:21:38.588Z",
"dateUpdated": "2026-11-16T07:47:48.690Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.2"
}
CVE-2025-3365 (GCVE-0-2025-3365)
Vulnerability from cvelistv5 – Published: 2025-06-06 08:14 – Updated: 2025-06-06 17:12
VLAI
EPSS
VEX
Title
Relative Path Traversal in OnlineSuite
Summary
A missing protection against path traversal allows to access
any file on the server.
Severity
9.8 (Critical)
SSVC
Exploitation: none
Automatable: yes
Technical Impact: total
CISA Coordinator · CISA-ADP (v2.0.3)
Decision recorded 2025-06-06 17:03 UTC
CWE
- CWE-23 - Relative Path Traversal
Assigner
References
1 reference
Impacted products
1 product
| Vendor | Product | Version | |
|---|---|---|---|
| B. Braun Melsungen AG | OnlineSuite |
Affected:
3.0
(custom)
|
Date Public
2025-06-06 07:00
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2025-3365",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "yes"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-06-06T17:03:10.577417Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-06-06T17:12:51.401Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"product": "OnlineSuite",
"vendor": "B. Braun Melsungen AG",
"versions": [
{
"status": "affected",
"version": "3.0",
"versionType": "custom"
}
]
}
],
"credits": [
{
"lang": "en",
"type": "reporter",
"value": "Fabian Weber (CODE WHITE GmbH)"
},
{
"lang": "en",
"type": "reporter",
"value": "Dr. Florian Hauser (CODE WHITE GmbH)"
}
],
"datePublic": "2025-06-06T07:00:00.000Z",
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003eA missing protection against path traversal allows to access\nany file on the server.\u003c/p\u003e"
}
],
"value": "A missing protection against path traversal allows to access\nany file on the server."
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 9.8,
"baseSeverity": "CRITICAL",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-23",
"description": "CWE-23 Relative Path Traversal",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2025-06-06T08:14:00.444Z",
"orgId": "653264ec-f98b-4e8f-b8b4-540a01b7657d",
"shortName": "B.Braun"
},
"references": [
{
"url": "https://www.bbraun.com/productsecurity"
}
],
"solutions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Field Service Information FSI 14-25 \u201cOnlineSuite AP3.0 - Security Fix\u201d provides a patch to these issues.\n\n\u003cbr\u003e"
}
],
"value": "Field Service Information FSI 14-25 \u201cOnlineSuite AP3.0 - Security Fix\u201d provides a patch to these issues."
}
],
"source": {
"discovery": "UNKNOWN"
},
"title": "Relative Path Traversal in OnlineSuite",
"x_generator": {
"engine": "Vulnogram 0.2.0"
}
}
},
"cveMetadata": {
"assignerOrgId": "653264ec-f98b-4e8f-b8b4-540a01b7657d",
"assignerShortName": "B.Braun",
"cveId": "CVE-2025-3365",
"datePublished": "2025-06-06T08:14:00.444Z",
"dateReserved": "2025-04-07T06:11:11.032Z",
"dateUpdated": "2025-06-06T17:12:51.401Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2025-3322 (GCVE-0-2025-3322)
Vulnerability from cvelistv5 – Published: 2025-06-06 08:13 – Updated: 2025-06-06 17:29
VLAI
EPSS
VEX
Title
Improper Neutralization of Special Elements in OnlineSuite
Summary
An improper neutralization of inputs used in expression
language allows remote code execution with the highest privileges on the
server.
Severity
SSVC
Exploitation: none
Automatable: yes
Technical Impact: total
CISA Coordinator · CISA-ADP (v2.0.3)
Decision recorded 2025-06-06 17:19 UTC
CWE
- CWE-917 - Improper Neutralization of Special Elements used in an Expression Language Statement ('Expression Language Injection')
Assigner
References
1 reference
Impacted products
1 product
| Vendor | Product | Version | |
|---|---|---|---|
| B. Braun Melsungen AG | OnlineSuite |
Affected:
3.0
(custom)
|
Date Public
2025-06-06 07:00
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2025-3322",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "yes"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-06-06T17:19:28.552605Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-06-06T17:29:30.972Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"product": "OnlineSuite",
"vendor": "B. Braun Melsungen AG",
"versions": [
{
"status": "affected",
"version": "3.0",
"versionType": "custom"
}
]
}
],
"credits": [
{
"lang": "en",
"type": "reporter",
"value": "Fabian Weber (CODE WHITE GmbH)"
},
{
"lang": "en",
"type": "reporter",
"value": "Dr. Florian Hauser (CODE WHITE GmbH)"
}
],
"datePublic": "2025-06-06T07:00:00.000Z",
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003eAn improper neutralization of inputs used in expression\nlanguage allows remote code execution with the highest privileges on the\nserver.\u003c/p\u003e"
}
],
"value": "An improper neutralization of inputs used in expression\nlanguage allows remote code execution with the highest privileges on the\nserver."
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "NONE",
"attackVector": "NETWORK",
"baseScore": 10,
"baseSeverity": "CRITICAL",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "HIGH",
"subConfidentialityImpact": "HIGH",
"subIntegrityImpact": "HIGH",
"userInteraction": "NONE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:H/SI:H/SA:H",
"version": "4.0",
"vulnAvailabilityImpact": "HIGH",
"vulnConfidentialityImpact": "HIGH",
"vulnIntegrityImpact": "HIGH",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-917",
"description": "CWE-917 Improper Neutralization of Special Elements used in an Expression Language Statement (\u0027Expression Language Injection\u0027)",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2025-06-06T08:13:12.028Z",
"orgId": "653264ec-f98b-4e8f-b8b4-540a01b7657d",
"shortName": "B.Braun"
},
"references": [
{
"url": "https://www.bbraun.com/productsecurity"
}
],
"solutions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Field Service Information FSI 14-25 \u201cOnlineSuite AP3.0 - Security Fix\u201d provides a patch to these issues.\n\n\u003cbr\u003e"
}
],
"value": "Field Service Information FSI 14-25 \u201cOnlineSuite AP3.0 - Security Fix\u201d provides a patch to these issues."
}
],
"source": {
"discovery": "UNKNOWN"
},
"title": "Improper Neutralization of Special Elements in OnlineSuite",
"x_generator": {
"engine": "Vulnogram 0.2.0"
}
}
},
"cveMetadata": {
"assignerOrgId": "653264ec-f98b-4e8f-b8b4-540a01b7657d",
"assignerShortName": "B.Braun",
"cveId": "CVE-2025-3322",
"datePublished": "2025-06-06T08:13:12.028Z",
"dateReserved": "2025-04-05T19:02:30.304Z",
"dateUpdated": "2025-06-06T17:29:30.972Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2025-3321 (GCVE-0-2025-3321)
Vulnerability from cvelistv5 – Published: 2025-06-06 08:12 – Updated: 2025-06-06 18:25
VLAI
EPSS
VEX
Title
Use of Hard-coded Credentials in OnlineSuite
Summary
A predefined administrative account is not documented and cannot
be deactivated. This account cannot be misused from the network, only by local
users on the server.
Severity
SSVC
Exploitation: none
Automatable: no
Technical Impact: total
CISA Coordinator · CISA-ADP (v2.0.3)
Decision recorded 2025-06-06 17:42 UTC
CWE
- CWE-798 - Use of Hard-coded Credentials
Assigner
References
1 reference
Impacted products
1 product
| Vendor | Product | Version | |
|---|---|---|---|
| B. Braun Melsungen AG | OnlineSuite |
Affected:
3.0
(custom)
|
Date Public
2025-06-06 07:00
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2025-3321",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-06-06T17:42:18.841236Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-06-06T18:25:54.094Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"product": "OnlineSuite",
"vendor": "B. Braun Melsungen AG",
"versions": [
{
"status": "affected",
"version": "3.0",
"versionType": "custom"
}
]
}
],
"credits": [
{
"lang": "en",
"type": "reporter",
"value": "Fabian Weber (CODE WHITE GmbH)"
},
{
"lang": "en",
"type": "reporter",
"value": "Dr. Florian Hauser (CODE WHITE GmbH)"
}
],
"datePublic": "2025-06-06T07:00:00.000Z",
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003e\u003c/p\u003e\u003cp\u003eA predefined administrative account is not documented and cannot\nbe deactivated. This account cannot be misused from the network, only by local\nusers on the server.\u003c/p\u003e\n\n\n\n\n\n\u003cp\u003e\u003c/p\u003e"
}
],
"value": "A predefined administrative account is not documented and cannot\nbe deactivated. This account cannot be misused from the network, only by local\nusers on the server."
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "NONE",
"attackVector": "LOCAL",
"baseScore": 9.4,
"baseSeverity": "CRITICAL",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "HIGH",
"subConfidentialityImpact": "HIGH",
"subIntegrityImpact": "HIGH",
"userInteraction": "NONE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:H/SI:H/SA:H",
"version": "4.0",
"vulnAvailabilityImpact": "HIGH",
"vulnConfidentialityImpact": "HIGH",
"vulnIntegrityImpact": "HIGH",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-798",
"description": "CWE-798 Use of Hard-coded Credentials",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2025-06-06T08:12:46.971Z",
"orgId": "653264ec-f98b-4e8f-b8b4-540a01b7657d",
"shortName": "B.Braun"
},
"references": [
{
"url": "https://www.bbraun.com/productsecurity"
}
],
"solutions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Field Service Information FSI 14-25 \u201cOnlineSuite AP3.0 - Security Fix\u201d provides a patch to these issues."
}
],
"value": "Field Service Information FSI 14-25 \u201cOnlineSuite AP3.0 - Security Fix\u201d provides a patch to these issues."
}
],
"source": {
"discovery": "UNKNOWN"
},
"title": "Use of Hard-coded Credentials in OnlineSuite",
"x_generator": {
"engine": "Vulnogram 0.2.0"
}
}
},
"cveMetadata": {
"assignerOrgId": "653264ec-f98b-4e8f-b8b4-540a01b7657d",
"assignerShortName": "B.Braun",
"cveId": "CVE-2025-3321",
"datePublished": "2025-06-06T08:12:46.971Z",
"dateReserved": "2025-04-05T19:01:47.895Z",
"dateUpdated": "2025-06-06T18:25:54.094Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2023-0888 (GCVE-0-2023-0888)
Vulnerability from cvelistv5 – Published: 2023-03-13 08:14 – Updated: 2025-02-27 18:17
VLAI
EPSS
VEX
Title
Authenticated eval injection in B. Braun Space Battery pack SP with Wi-Fi
Summary
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
Severity
4.9 (Medium)
SSVC
Exploitation: none
Automatable: no
Technical Impact: partial
CISA Coordinator · CISA-ADP (v2.0.3)
Decision recorded 2025-02-27 18:17 UTC
CWE
- CWE-95 - Eval injection
Assigner
References
2 references
Impacted products
1 product
| Vendor | Product | Version | |
|---|---|---|---|
| B. Braun | Space Battery Pack SP with Wi-Fi |
Affected:
0 , ≤ 053L000092
(custom)
Affected: 0 , ≤ 054U000092 (custom) |
Date Public
2023-03-15 07:00
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-02T05:24:34.770Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_transferred"
],
"url": "https://www.bbraun.com/productsecurity"
},
{
"tags": [
"x_transferred"
],
"url": "https://www.bbraunusa.com/productsecurity"
}
],
"title": "CVE Program Container"
},
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2023-0888",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2025-02-27T18:17:34.759625Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2025-02-27T18:17:42.996Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"product": "Space Battery Pack SP with Wi-Fi",
"vendor": "B. Braun ",
"versions": [
{
"lessThanOrEqual": "053L000092",
"status": "affected",
"version": "0",
"versionType": "custom"
},
{
"lessThanOrEqual": "054U000092",
"status": "affected",
"version": "0",
"versionType": "custom"
}
]
}
],
"credits": [
{
"lang": "en",
"type": "finder",
"user": "00000000-0000-4000-9000-000000000000",
"value": "Tom Johnston"
}
],
"datePublic": "2023-03-15T07:00:00.000Z",
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks\u003cbr\u003e\u003cbr\u003e"
}
],
"value": "An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks\n\n"
}
],
"impacts": [
{
"capecId": "CAPEC-233",
"descriptions": [
{
"lang": "en",
"value": "CAPEC-233 Privilege Escalation"
}
]
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "NETWORK",
"availabilityImpact": "NONE",
"baseScore": 4.9,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "NONE",
"integrityImpact": "HIGH",
"privilegesRequired": "HIGH",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:N/I:H/A:N",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-95",
"description": "CWE-95 Eval injection",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2023-03-13T08:14:32.358Z",
"orgId": "653264ec-f98b-4e8f-b8b4-540a01b7657d",
"shortName": "B.Braun"
},
"references": [
{
"url": "https://www.bbraun.com/productsecurity"
},
{
"url": "https://www.bbraunusa.com/productsecurity"
}
],
"solutions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Update to version L93/U93 or newer solves the issue."
}
],
"value": "Update to version L93/U93 or newer solves the issue."
}
],
"source": {
"discovery": "EXTERNAL"
},
"title": "Authenticated eval injection in B. Braun Space Battery pack SP with Wi-Fi ",
"workarounds": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Use secure credentials and keep them secret.\u0026nbsp;\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eSecure your network to prevent outside access to medical devices.\u003c/span\u003e"
}
],
"value": "Use secure credentials and keep them secret.\u00a0Secure your network to prevent outside access to medical devices."
}
],
"x_generator": {
"engine": "Vulnogram 0.1.0-dev"
}
}
},
"cveMetadata": {
"assignerOrgId": "653264ec-f98b-4e8f-b8b4-540a01b7657d",
"assignerShortName": "B.Braun",
"cveId": "CVE-2023-0888",
"datePublished": "2023-03-13T08:14:32.358Z",
"dateReserved": "2023-02-17T09:22:21.832Z",
"dateUpdated": "2025-02-27T18:17:42.996Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}