Search

Find a vulnerability

Search criteria

    4 vulnerabilities found for Infinity Delta by Dräger

    CVE-2019-25717 (GCVE-0-2019-25717)

    Vulnerability from nvd – Published: 2026-06-02 13:42 – Updated: 2026-07-15 01:24
    VLAI
    Title
    Dräger Infinity Delta/Kappa Patient Monitors Unauthenticated Log File Disclosure
    Summary
    Dräger Infinity Delta, Delta XL, and Kappa patient monitors contain an information disclosure vulnerability that allows unauthenticated network attackers to access log files over a network connection. Attackers can retrieve device internals, location information, and wired network configuration details from the exposed log files.
    SSVC
    Exploitation: none Automatable: no Technical Impact: partial
    CISA Coordinator · CISA-ADP (v2.0.3)
    Decision recorded 2026-06-03 13:45 UTC
    CWE
    • CWE-538 - Insertion of Sensitive Information into Externally-Accessible File or Directory
    References
    Impacted products
    Vendor Product Version
    Dräger Infinity Delta Affected: all software versions (custom)
    Create a notification for this product.
    Dräger Infinity Delta XL Affected: all software versions (custom)
    Create a notification for this product.
    Dräger Infinity Kappa Affected: all software versions (custom)
    Create a notification for this product.
    Date Public
    2019-01-22 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2019-25717",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "partial"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2026-06-03T13:45:52.446853Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2026-06-03T13:46:03.638Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unaffected",
              "product": "Infinity Delta",
              "vendor": "Dr\u00e4ger",
              "versions": [
                {
                  "status": "affected",
                  "version": "all software versions",
                  "versionType": "custom"
                }
              ]
            },
            {
              "defaultStatus": "unaffected",
              "product": "Infinity Delta XL",
              "vendor": "Dr\u00e4ger",
              "versions": [
                {
                  "status": "affected",
                  "version": "all software versions",
                  "versionType": "custom"
                }
              ]
            },
            {
              "defaultStatus": "unaffected",
              "product": "Infinity Kappa",
              "vendor": "Dr\u00e4ger",
              "versions": [
                {
                  "status": "affected",
                  "version": "all software versions",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "cpeApplicability": [
            {
              "nodes": [
                {
                  "cpeMatch": [
                    {
                      "criteria": "cpe:2.3:h:draeger:infinity_delta:all_software_versions:*:*:*:*:*:*:*",
                      "vulnerable": true
                    }
                  ],
                  "negate": false,
                  "operator": "OR"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "value": "Marc Ruef and Rocco Gagliardi, scip AG"
            }
          ],
          "datePublic": "2019-01-22T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003eDr\u00e4ger Infinity Delta, Delta XL, and Kappa patient monitors contain an information disclosure vulnerability that allows unauthenticated network attackers to access log files over a network connection. Attackers can retrieve device internals, location information, and wired network configuration details from the exposed log files.\u003c/p\u003e"
                }
              ],
              "value": "Dr\u00e4ger Infinity Delta, Delta XL, and Kappa patient monitors contain an information disclosure vulnerability that allows unauthenticated network attackers to access log files over a network connection. Attackers can retrieve device internals, location information, and wired network configuration details from the exposed log files."
            }
          ],
          "metrics": [
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "LOW",
                "attackRequirements": "NONE",
                "attackVector": "ADJACENT",
                "baseScore": 5.3,
                "baseSeverity": "MEDIUM",
                "exploitMaturity": "NOT_DEFINED",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "NONE",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:L/VI:N/VA:N/SC:N/SI:N/SA:N",
                "version": "4.0",
                "vulnAvailabilityImpact": "NONE",
                "vulnConfidentialityImpact": "LOW",
                "vulnIntegrityImpact": "NONE",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV3_1": {
                "attackComplexity": "LOW",
                "attackVector": "ADJACENT_NETWORK",
                "availabilityImpact": "NONE",
                "baseScore": 4.3,
                "baseSeverity": "MEDIUM",
                "confidentialityImpact": "LOW",
                "integrityImpact": "NONE",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-538",
                  "description": "CWE-538 Insertion of Sensitive Information into Externally-Accessible File or Directory",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-07-15T01:24:11.103Z",
            "orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
            "shortName": "VulnCheck"
          },
          "references": [
            {
              "tags": [
                "vendor-advisory"
              ],
              "url": "https://static.draeger.com/security"
            },
            {
              "tags": [
                "third-party-advisory"
              ],
              "url": "https://www.vulncheck.com/advisories/dr-ger-infinity-delta-kappa-patient-monitors-unauthenticated-log-file-disclosure"
            }
          ],
          "source": {
            "discovery": "EXTERNAL"
          },
          "title": "Dr\u00e4ger Infinity Delta/Kappa Patient Monitors Unauthenticated Log File Disclosure"
        }
      },
      "cveMetadata": {
        "assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
        "assignerShortName": "VulnCheck",
        "cveId": "CVE-2019-25717",
        "datePublished": "2026-06-02T13:42:35.033Z",
        "dateReserved": "2026-06-01T21:15:41.689Z",
        "dateUpdated": "2026-07-15T01:24:11.103Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }

    CVE-2019-25716 (GCVE-0-2019-25716)

    Vulnerability from nvd – Published: 2026-06-01 21:15 – Updated: 2026-07-15 01:24
    VLAI
    Title
    Dräger Infinity Delta/Kappa Patient Monitor DoS via Malformed Network Packet
    Summary
    Dräger Infinity Delta, Delta XL, and Kappa patient monitors contain a denial-of-service vulnerability that allows remote attackers to cause the monitor to reboot by sending a malformed network packet. Attackers can repeatedly send malformed network packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.
    SSVC
    Exploitation: none Automatable: no Technical Impact: partial
    CISA Coordinator · CISA-ADP (v2.0.3)
    Decision recorded 2026-06-02 15:09 UTC
    CWE
    • CWE-15 - External Control of System or Configuration Setting
    References
    Impacted products
    Vendor Product Version
    Dräger Infinity Delta Affected: Infinity Delta (custom)
    Create a notification for this product.
    Dräger Infinity Delta XL Affected: Infinity Delta XL (custom)
    Create a notification for this product.
    Dräger Infinity Kappa Affected: Infinity Kappa (custom)
    Create a notification for this product.
    Date Public
    2019-01-22 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2019-25716",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "partial"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2026-06-02T15:09:36.384627Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2026-06-02T15:45:56.330Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unaffected",
              "product": "Infinity Delta",
              "vendor": "Dr\u00e4ger",
              "versions": [
                {
                  "status": "affected",
                  "version": "Infinity Delta",
                  "versionType": "custom"
                }
              ]
            },
            {
              "defaultStatus": "unaffected",
              "product": "Infinity Delta XL",
              "vendor": "Dr\u00e4ger",
              "versions": [
                {
                  "status": "affected",
                  "version": "Infinity Delta XL",
                  "versionType": "custom"
                }
              ]
            },
            {
              "defaultStatus": "unaffected",
              "product": "Infinity Kappa",
              "vendor": "Dr\u00e4ger",
              "versions": [
                {
                  "status": "affected",
                  "version": "Infinity Kappa",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "cpeApplicability": [
            {
              "nodes": [
                {
                  "cpeMatch": [
                    {
                      "criteria": "cpe:2.3:h:draeger:infinity_delta:infinity_delta:*:*:*:*:*:*:*",
                      "vulnerable": true
                    },
                    {
                      "criteria": "cpe:2.3:h:draeger:infinity_delta:infinity_delta_xl:*:*:*:*:*:*:*",
                      "vulnerable": true
                    }
                  ],
                  "negate": false,
                  "operator": "OR"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "value": "Marc Ruef and Rocco Gagliardi, scip AG"
            }
          ],
          "datePublic": "2019-01-22T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003eDr\u00e4ger Infinity Delta, Delta XL, and Kappa patient monitors contain a denial-of-service vulnerability that allows remote attackers to cause the monitor to reboot by sending a malformed network packet. Attackers can repeatedly send malformed network packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.\u003c/p\u003e"
                }
              ],
              "value": "Dr\u00e4ger Infinity Delta, Delta XL, and Kappa patient monitors contain a denial-of-service vulnerability that allows remote attackers to cause the monitor to reboot by sending a malformed network packet. Attackers can repeatedly send malformed network packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity."
            }
          ],
          "metrics": [
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "LOW",
                "attackRequirements": "NONE",
                "attackVector": "ADJACENT",
                "baseScore": 7.1,
                "baseSeverity": "HIGH",
                "exploitMaturity": "NOT_DEFINED",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "NONE",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N",
                "version": "4.0",
                "vulnAvailabilityImpact": "HIGH",
                "vulnConfidentialityImpact": "NONE",
                "vulnIntegrityImpact": "NONE",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV3_1": {
                "attackComplexity": "LOW",
                "attackVector": "ADJACENT_NETWORK",
                "availabilityImpact": "HIGH",
                "baseScore": 6.5,
                "baseSeverity": "MEDIUM",
                "confidentialityImpact": "NONE",
                "integrityImpact": "NONE",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-15",
                  "description": "CWE-15 External Control of System or Configuration Setting",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-07-15T01:24:10.429Z",
            "orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
            "shortName": "VulnCheck"
          },
          "references": [
            {
              "tags": [
                "vendor-advisory"
              ],
              "url": "https://static.draeger.com/security/download/2019-01-22-draeger-infinity-delta-vf10-1-security-advisory.pdf"
            },
            {
              "tags": [
                "third-party-advisory"
              ],
              "url": "https://www.vulncheck.com/advisories/dr-ger-infinity-delta-kappa-patient-monitor-dos-via-malformed-network-packet"
            }
          ],
          "source": {
            "discovery": "EXTERNAL"
          },
          "title": "Dr\u00e4ger Infinity Delta/Kappa Patient Monitor DoS via Malformed Network Packet",
          "x_generator": {
            "engine": "Vulnogram 1.0.2"
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
        "assignerShortName": "VulnCheck",
        "cveId": "CVE-2019-25716",
        "datePublished": "2026-06-01T21:15:07.156Z",
        "dateReserved": "2026-06-01T20:44:47.913Z",
        "dateUpdated": "2026-07-15T01:24:10.429Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }

    CVE-2019-25717 (GCVE-0-2019-25717)

    Vulnerability from cvelistv5 – Published: 2026-06-02 13:42 – Updated: 2026-07-15 01:24
    VLAI
    Title
    Dräger Infinity Delta/Kappa Patient Monitors Unauthenticated Log File Disclosure
    Summary
    Dräger Infinity Delta, Delta XL, and Kappa patient monitors contain an information disclosure vulnerability that allows unauthenticated network attackers to access log files over a network connection. Attackers can retrieve device internals, location information, and wired network configuration details from the exposed log files.
    SSVC
    Exploitation: none Automatable: no Technical Impact: partial
    CISA Coordinator · CISA-ADP (v2.0.3)
    Decision recorded 2026-06-03 13:45 UTC
    CWE
    • CWE-538 - Insertion of Sensitive Information into Externally-Accessible File or Directory
    References
    Impacted products
    Vendor Product Version
    Dräger Infinity Delta Affected: all software versions (custom)
    Create a notification for this product.
    Dräger Infinity Delta XL Affected: all software versions (custom)
    Create a notification for this product.
    Dräger Infinity Kappa Affected: all software versions (custom)
    Create a notification for this product.
    Date Public
    2019-01-22 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2019-25717",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "partial"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2026-06-03T13:45:52.446853Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2026-06-03T13:46:03.638Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unaffected",
              "product": "Infinity Delta",
              "vendor": "Dr\u00e4ger",
              "versions": [
                {
                  "status": "affected",
                  "version": "all software versions",
                  "versionType": "custom"
                }
              ]
            },
            {
              "defaultStatus": "unaffected",
              "product": "Infinity Delta XL",
              "vendor": "Dr\u00e4ger",
              "versions": [
                {
                  "status": "affected",
                  "version": "all software versions",
                  "versionType": "custom"
                }
              ]
            },
            {
              "defaultStatus": "unaffected",
              "product": "Infinity Kappa",
              "vendor": "Dr\u00e4ger",
              "versions": [
                {
                  "status": "affected",
                  "version": "all software versions",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "cpeApplicability": [
            {
              "nodes": [
                {
                  "cpeMatch": [
                    {
                      "criteria": "cpe:2.3:h:draeger:infinity_delta:all_software_versions:*:*:*:*:*:*:*",
                      "vulnerable": true
                    }
                  ],
                  "negate": false,
                  "operator": "OR"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "value": "Marc Ruef and Rocco Gagliardi, scip AG"
            }
          ],
          "datePublic": "2019-01-22T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003eDr\u00e4ger Infinity Delta, Delta XL, and Kappa patient monitors contain an information disclosure vulnerability that allows unauthenticated network attackers to access log files over a network connection. Attackers can retrieve device internals, location information, and wired network configuration details from the exposed log files.\u003c/p\u003e"
                }
              ],
              "value": "Dr\u00e4ger Infinity Delta, Delta XL, and Kappa patient monitors contain an information disclosure vulnerability that allows unauthenticated network attackers to access log files over a network connection. Attackers can retrieve device internals, location information, and wired network configuration details from the exposed log files."
            }
          ],
          "metrics": [
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "LOW",
                "attackRequirements": "NONE",
                "attackVector": "ADJACENT",
                "baseScore": 5.3,
                "baseSeverity": "MEDIUM",
                "exploitMaturity": "NOT_DEFINED",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "NONE",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:L/VI:N/VA:N/SC:N/SI:N/SA:N",
                "version": "4.0",
                "vulnAvailabilityImpact": "NONE",
                "vulnConfidentialityImpact": "LOW",
                "vulnIntegrityImpact": "NONE",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV3_1": {
                "attackComplexity": "LOW",
                "attackVector": "ADJACENT_NETWORK",
                "availabilityImpact": "NONE",
                "baseScore": 4.3,
                "baseSeverity": "MEDIUM",
                "confidentialityImpact": "LOW",
                "integrityImpact": "NONE",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-538",
                  "description": "CWE-538 Insertion of Sensitive Information into Externally-Accessible File or Directory",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-07-15T01:24:11.103Z",
            "orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
            "shortName": "VulnCheck"
          },
          "references": [
            {
              "tags": [
                "vendor-advisory"
              ],
              "url": "https://static.draeger.com/security"
            },
            {
              "tags": [
                "third-party-advisory"
              ],
              "url": "https://www.vulncheck.com/advisories/dr-ger-infinity-delta-kappa-patient-monitors-unauthenticated-log-file-disclosure"
            }
          ],
          "source": {
            "discovery": "EXTERNAL"
          },
          "title": "Dr\u00e4ger Infinity Delta/Kappa Patient Monitors Unauthenticated Log File Disclosure"
        }
      },
      "cveMetadata": {
        "assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
        "assignerShortName": "VulnCheck",
        "cveId": "CVE-2019-25717",
        "datePublished": "2026-06-02T13:42:35.033Z",
        "dateReserved": "2026-06-01T21:15:41.689Z",
        "dateUpdated": "2026-07-15T01:24:11.103Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }

    CVE-2019-25716 (GCVE-0-2019-25716)

    Vulnerability from cvelistv5 – Published: 2026-06-01 21:15 – Updated: 2026-07-15 01:24
    VLAI
    Title
    Dräger Infinity Delta/Kappa Patient Monitor DoS via Malformed Network Packet
    Summary
    Dräger Infinity Delta, Delta XL, and Kappa patient monitors contain a denial-of-service vulnerability that allows remote attackers to cause the monitor to reboot by sending a malformed network packet. Attackers can repeatedly send malformed network packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.
    SSVC
    Exploitation: none Automatable: no Technical Impact: partial
    CISA Coordinator · CISA-ADP (v2.0.3)
    Decision recorded 2026-06-02 15:09 UTC
    CWE
    • CWE-15 - External Control of System or Configuration Setting
    References
    Impacted products
    Vendor Product Version
    Dräger Infinity Delta Affected: Infinity Delta (custom)
    Create a notification for this product.
    Dräger Infinity Delta XL Affected: Infinity Delta XL (custom)
    Create a notification for this product.
    Dräger Infinity Kappa Affected: Infinity Kappa (custom)
    Create a notification for this product.
    Date Public
    2019-01-22 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2019-25716",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "partial"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2026-06-02T15:09:36.384627Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2026-06-02T15:45:56.330Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unaffected",
              "product": "Infinity Delta",
              "vendor": "Dr\u00e4ger",
              "versions": [
                {
                  "status": "affected",
                  "version": "Infinity Delta",
                  "versionType": "custom"
                }
              ]
            },
            {
              "defaultStatus": "unaffected",
              "product": "Infinity Delta XL",
              "vendor": "Dr\u00e4ger",
              "versions": [
                {
                  "status": "affected",
                  "version": "Infinity Delta XL",
                  "versionType": "custom"
                }
              ]
            },
            {
              "defaultStatus": "unaffected",
              "product": "Infinity Kappa",
              "vendor": "Dr\u00e4ger",
              "versions": [
                {
                  "status": "affected",
                  "version": "Infinity Kappa",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "cpeApplicability": [
            {
              "nodes": [
                {
                  "cpeMatch": [
                    {
                      "criteria": "cpe:2.3:h:draeger:infinity_delta:infinity_delta:*:*:*:*:*:*:*",
                      "vulnerable": true
                    },
                    {
                      "criteria": "cpe:2.3:h:draeger:infinity_delta:infinity_delta_xl:*:*:*:*:*:*:*",
                      "vulnerable": true
                    }
                  ],
                  "negate": false,
                  "operator": "OR"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "value": "Marc Ruef and Rocco Gagliardi, scip AG"
            }
          ],
          "datePublic": "2019-01-22T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003eDr\u00e4ger Infinity Delta, Delta XL, and Kappa patient monitors contain a denial-of-service vulnerability that allows remote attackers to cause the monitor to reboot by sending a malformed network packet. Attackers can repeatedly send malformed network packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.\u003c/p\u003e"
                }
              ],
              "value": "Dr\u00e4ger Infinity Delta, Delta XL, and Kappa patient monitors contain a denial-of-service vulnerability that allows remote attackers to cause the monitor to reboot by sending a malformed network packet. Attackers can repeatedly send malformed network packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity."
            }
          ],
          "metrics": [
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "LOW",
                "attackRequirements": "NONE",
                "attackVector": "ADJACENT",
                "baseScore": 7.1,
                "baseSeverity": "HIGH",
                "exploitMaturity": "NOT_DEFINED",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "NONE",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N",
                "version": "4.0",
                "vulnAvailabilityImpact": "HIGH",
                "vulnConfidentialityImpact": "NONE",
                "vulnIntegrityImpact": "NONE",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV3_1": {
                "attackComplexity": "LOW",
                "attackVector": "ADJACENT_NETWORK",
                "availabilityImpact": "HIGH",
                "baseScore": 6.5,
                "baseSeverity": "MEDIUM",
                "confidentialityImpact": "NONE",
                "integrityImpact": "NONE",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-15",
                  "description": "CWE-15 External Control of System or Configuration Setting",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-07-15T01:24:10.429Z",
            "orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
            "shortName": "VulnCheck"
          },
          "references": [
            {
              "tags": [
                "vendor-advisory"
              ],
              "url": "https://static.draeger.com/security/download/2019-01-22-draeger-infinity-delta-vf10-1-security-advisory.pdf"
            },
            {
              "tags": [
                "third-party-advisory"
              ],
              "url": "https://www.vulncheck.com/advisories/dr-ger-infinity-delta-kappa-patient-monitor-dos-via-malformed-network-packet"
            }
          ],
          "source": {
            "discovery": "EXTERNAL"
          },
          "title": "Dr\u00e4ger Infinity Delta/Kappa Patient Monitor DoS via Malformed Network Packet",
          "x_generator": {
            "engine": "Vulnogram 1.0.2"
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
        "assignerShortName": "VulnCheck",
        "cveId": "CVE-2019-25716",
        "datePublished": "2026-06-01T21:15:07.156Z",
        "dateReserved": "2026-06-01T20:44:47.913Z",
        "dateUpdated": "2026-07-15T01:24:10.429Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }