    <item>
      <title>CVE-2025-14096 — Credential Disclosure vulnerability in Radiometer Products</title>
      <link>https://cve.radiocsirt.org/vuln/cve-2025-14096</link>
      <description>&lt;p&gt;&lt;strong&gt;Affected:&lt;/strong&gt; Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers, Radiometer Medical Aps AQT90 FLEX Analyzers, Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX analyzers&lt;/p&gt;
&lt;p&gt;A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.&lt;/p&gt;
&lt;p&gt;Other related CVE&amp;#39;s are CVE-2025-14095 &amp;amp; CVE-2025-14097.&lt;/p&gt;
&lt;p&gt;Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.&lt;/p&gt;
&lt;p&gt;Required Configuration for Exposure:&lt;/p&gt;
&lt;p&gt;Attacker requires physical access to the analyzer.&lt;/p&gt;
&lt;p&gt;Temporary work Around:
Only authorized people can physically access the analyzer.&lt;/p&gt;
&lt;p&gt;Permanent solution:
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.&lt;/p&gt;
&lt;p&gt;Exploit Status:&lt;/p&gt;
&lt;p&gt;Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.&lt;/p&gt;</description>
      <content:encoded>&lt;p&gt;&lt;strong&gt;Affected:&lt;/strong&gt; Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers, Radiometer Medical Aps AQT90 FLEX Analyzers, Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX analyzers&lt;/p&gt;
&lt;p&gt;A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.&lt;/p&gt;
&lt;p&gt;Other related CVE&amp;#39;s are CVE-2025-14095 &amp;amp; CVE-2025-14097.&lt;/p&gt;
&lt;p&gt;Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.&lt;/p&gt;
&lt;p&gt;Required Configuration for Exposure:&lt;/p&gt;
&lt;p&gt;Attacker requires physical access to the analyzer.&lt;/p&gt;
&lt;p&gt;Temporary work Around:
Only authorized people can physically access the analyzer.&lt;/p&gt;
&lt;p&gt;Permanent solution:
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.&lt;/p&gt;
&lt;p&gt;Exploit Status:&lt;/p&gt;
&lt;p&gt;Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.&lt;/p&gt;</content:encoded>
      <guid isPermaLink="false">https://cve.radiocsirt.org/vuln/cve-2025-14096</guid>