<?xml version='1.0' encoding='UTF-8'?>
<?xml-stylesheet href="/static/style.xsl" type="text/xsl"?>
<rss xmlns:atom="http://www.w3.org/2005/Atom" xmlns:content="http://purl.org/rss/1.0/modules/content/" version="2.0">
  <channel>
    <title>Most recent entries from all</title>
    <link>https://cve.radiocsirt.org</link>
    <description>Contains only the most 10 recent entries.</description>
    <docs>http://www.rssboard.org/rss-specification</docs>
    <generator>python-feedgen</generator>
    <language>en</language>
    <lastBuildDate>Tue, 06 Oct 2026 21:28:38 +0000</lastBuildDate>
    <item>
      <title>EUVD-2026-263825</title>
      <link>https://cve.radiocsirt.org/vuln/euvd-2026-263825</link>
      <description>EUVD-2026-263825</description>
      <guid isPermaLink="false">https://cve.radiocsirt.org/vuln/euvd-2026-263825</guid>
    </item>
    <item>
      <title>fkie_cve-2025-14096</title>
      <link>https://cve.radiocsirt.org/vuln/fkie_cve-2025-14096</link>
      <description>&lt;p&gt;A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.&lt;/p&gt;
&lt;p&gt;Other related CVE&amp;#39;s are CVE-2025-14095 &amp;amp; CVE-2025-14097.&lt;/p&gt;
&lt;p&gt;Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.&lt;/p&gt;
&lt;p&gt;Required Configuration for Exposure:&lt;/p&gt;
&lt;p&gt;Attacker requires physical access to the analyzer.&lt;/p&gt;
&lt;p&gt;Temporary work Around:
Only authorized people can physically access the analyzer.&lt;/p&gt;
&lt;p&gt;Permanent solution:
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.&lt;/p&gt;
&lt;p&gt;Exploit Status:&lt;/p&gt;
&lt;p&gt;Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.&lt;/p&gt;</description>
      <content:encoded>&lt;p&gt;A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.&lt;/p&gt;
&lt;p&gt;Other related CVE&amp;#39;s are CVE-2025-14095 &amp;amp; CVE-2025-14097.&lt;/p&gt;
&lt;p&gt;Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.&lt;/p&gt;
&lt;p&gt;Required Configuration for Exposure:&lt;/p&gt;
&lt;p&gt;Attacker requires physical access to the analyzer.&lt;/p&gt;
&lt;p&gt;Temporary work Around:
Only authorized people can physically access the analyzer.&lt;/p&gt;
&lt;p&gt;Permanent solution:
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.&lt;/p&gt;
&lt;p&gt;Exploit Status:&lt;/p&gt;
&lt;p&gt;Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.&lt;/p&gt;</content:encoded>
      <guid isPermaLink="false">https://cve.radiocsirt.org/vuln/fkie_cve-2025-14096</guid>
    </item>
    <item>
      <title>GHSA-h5qw-42jq-7cmh</title>
      <link>https://cve.radiocsirt.org/vuln/ghsa-h5qw-42jq-7cmh</link>
      <description>&lt;p&gt;A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.&lt;/p&gt;
&lt;p&gt;Other related CVE&amp;#39;s are CVE-2025-14095 &amp;amp; CVE-2025-14097.&lt;/p&gt;
&lt;p&gt;Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.&lt;/p&gt;
&lt;p&gt;Required Configuration for Exposure:&lt;/p&gt;
&lt;p&gt;Attacker requires physical access to the analyzer.&lt;/p&gt;
&lt;p&gt;Temporary work Around:
Only authorized people can physically access the analyzer.&lt;/p&gt;
&lt;p&gt;Permanent solution:
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.&lt;/p&gt;
&lt;p&gt;Exploit Status:&lt;/p&gt;
&lt;p&gt;Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.&lt;/p&gt;</description>
      <content:encoded>&lt;p&gt;A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.&lt;/p&gt;
&lt;p&gt;Other related CVE&amp;#39;s are CVE-2025-14095 &amp;amp; CVE-2025-14097.&lt;/p&gt;
&lt;p&gt;Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.&lt;/p&gt;
&lt;p&gt;Required Configuration for Exposure:&lt;/p&gt;
&lt;p&gt;Attacker requires physical access to the analyzer.&lt;/p&gt;
&lt;p&gt;Temporary work Around:
Only authorized people can physically access the analyzer.&lt;/p&gt;
&lt;p&gt;Permanent solution:
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.&lt;/p&gt;
&lt;p&gt;Exploit Status:&lt;/p&gt;
&lt;p&gt;Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.&lt;/p&gt;</content:encoded>
      <guid isPermaLink="false">https://cve.radiocsirt.org/vuln/ghsa-h5qw-42jq-7cmh</guid>
    </item>
  </channel>
</rss>
