    <item>
      <title>CVE-2018-14781 — Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay</title>
      <link>https://cve.radiocsirt.org/vuln/cve-2018-14781</link>
      <description>&lt;p&gt;&lt;strong&gt;Affected:&lt;/strong&gt; Medtronic MMT- 508 - MiniMed pump, Medtronic MMT – 511 pump Paradigm, Medtronic MMT – 512 / MMT – 712 Paradigm x12, Medtronic MMT – 515 / MMT – 715 Paradigm x15, Medtronic MMT – 522 / MMT – 722 Paradigm REAL-TIME, Medtronic MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIME, Medtronic MMT – 523 / MMT – 723 Paradigm Revel, Medtronic MMT – 523(K) / MMT – 723(K) Paradigm, Medtronic MMT – 554 / MMT – 754 MiniMed Veo, Medtronic MMT – 551 / MMT – 751 MiniMed 530G&lt;/p&gt;
&lt;p&gt;Medtronic MiniMed MMT&lt;/p&gt;
&lt;p&gt;devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.&lt;/p&gt;</description>
      <content:encoded>&lt;p&gt;&lt;strong&gt;Affected:&lt;/strong&gt; Medtronic MMT- 508 - MiniMed pump, Medtronic MMT – 511 pump Paradigm, Medtronic MMT – 512 / MMT – 712 Paradigm x12, Medtronic MMT – 515 / MMT – 715 Paradigm x15, Medtronic MMT – 522 / MMT – 722 Paradigm REAL-TIME, Medtronic MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIME, Medtronic MMT – 523 / MMT – 723 Paradigm Revel, Medtronic MMT – 523(K) / MMT – 723(K) Paradigm, Medtronic MMT – 554 / MMT – 754 MiniMed Veo, Medtronic MMT – 551 / MMT – 751 MiniMed 530G&lt;/p&gt;
&lt;p&gt;Medtronic MiniMed MMT&lt;/p&gt;
&lt;p&gt;devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.&lt;/p&gt;</content:encoded>
      <guid isPermaLink="false">https://cve.radiocsirt.org/vuln/cve-2018-14781</guid>