<?xml version='1.0' encoding='UTF-8'?>
<?xml-stylesheet href="/static/style.xsl" type="text/xsl"?>
<rss xmlns:atom="http://www.w3.org/2005/Atom" xmlns:content="http://purl.org/rss/1.0/modules/content/" version="2.0">
  <channel>
    <title>Most recent entries from all</title>
    <link>https://cve.radiocsirt.org</link>
    <description>Contains only the most 10 recent entries.</description>
    <docs>http://www.rssboard.org/rss-specification</docs>
    <generator>python-feedgen</generator>
    <language>en</language>
    <lastBuildDate>Fri, 02 Oct 2026 20:51:58 +0000</lastBuildDate>
    <item>
      <title>EUVD-2026-201752</title>
      <link>https://cve.radiocsirt.org/vuln/euvd-2026-201752</link>
      <description>EUVD-2026-201752</description>
      <guid isPermaLink="false">https://cve.radiocsirt.org/vuln/euvd-2026-201752</guid>
    </item>
    <item>
      <title>fkie_cve-2024-48967</title>
      <link>https://cve.radiocsirt.org/vuln/fkie_cve-2024-48967</link>
      <description>&lt;p&gt;The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.&lt;/p&gt;</description>
      <content:encoded>&lt;p&gt;The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.&lt;/p&gt;</content:encoded>
      <guid isPermaLink="false">https://cve.radiocsirt.org/vuln/fkie_cve-2024-48967</guid>
    </item>
    <item>
      <title>GHSA-9mp2-83wh-h47x</title>
      <link>https://cve.radiocsirt.org/vuln/ghsa-9mp2-83wh-h47x</link>
      <description>&lt;p&gt;The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.&lt;/p&gt;</description>
      <content:encoded>&lt;p&gt;The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.&lt;/p&gt;</content:encoded>
      <guid isPermaLink="false">https://cve.radiocsirt.org/vuln/ghsa-9mp2-83wh-h47x</guid>
    </item>
    <item>
      <title>ICSMA-24-319-01 — Baxter Life2000 Ventilation System</title>
      <link>https://cve.radiocsirt.org/vuln/icsma-24-319-01</link>
      <description>&lt;p&gt;Improper data protection on the ventilator&amp;#39;s serial interface could allow an attacker to send and receive messages that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance. There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.. The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges. The debug port on the ventilator&amp;#39;s serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance. The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device&amp;#39;s configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the…&lt;/p&gt;</description>
      <content:encoded>&lt;p&gt;Improper data protection on the ventilator&amp;#39;s serial interface could allow an attacker to send and receive messages that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance. There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.. The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges. The debug port on the ventilator&amp;#39;s serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance. The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device&amp;#39;s configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the…&lt;/p&gt;</content:encoded>
      <guid isPermaLink="false">https://cve.radiocsirt.org/vuln/icsma-24-319-01</guid>
    </item>
  </channel>
</rss>