    <title>CVE-2022-36385 — Contec Health CMS8000</title>
    <updated>2026-10-05T21:08:47.304897+00:00</updated>
    <content type="xhtml">
      <div xmlns="http://www.w3.org/1999/xhtml"><p><strong>Affected:</strong> Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor</p>
<p>A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device.</p></div>