  <entry>
    <id>https://cve.radiocsirt.org/vuln/cve-2019-25716</id>
    <title>CVE-2019-25716 — Dräger Infinity Delta/Kappa Patient Monitor DoS via Malformed Network Packet</title>
    <updated>2026-10-07T01:19:02.006125+00:00</updated>
    <content type="xhtml">
      <div xmlns="http://www.w3.org/1999/xhtml"><p><strong>Affected:</strong> Dräger Infinity Delta, Dräger Infinity Delta XL, Dräger Infinity Kappa</p>
<p>Dräger Infinity Delta, Delta XL, and Kappa patient monitors contain a denial-of-service vulnerability that allows remote attackers to cause the monitor to reboot by sending a malformed network packet. Attackers can repeatedly send malformed network packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.</p></div>