<?xml version='1.0' encoding='UTF-8'?>
<?xml-stylesheet href="/static/style.xsl" type="text/xsl"?>
<feed xmlns="http://www.w3.org/2005/Atom" xml:lang="en">
  <id>https://cve.radiocsirt.org/rss/recent/all/10</id>
  <title>Most recent entries from all</title>
  <updated>2026-10-04T16:07:15.749692+00:00</updated>
  <author>
    <name>Vulnerability-Lookup</name>
    <email>csirt@opendfir.org</email>
  </author>
  <link href="https://cve.radiocsirt.org" rel="alternate"/>
  <generator uri="https://lkiesow.github.io/python-feedgen" version="1.0.0">python-feedgen</generator>
  <subtitle>Contains only the most 10 recent entries.</subtitle>
  <entry>
    <id>https://cve.radiocsirt.org/vuln/cnvd-2022-22468</id>
    <title>cnvd-2022-22468</title>
    <updated>2026-10-04T16:07:15.753482+00:00</updated>
    <content>cnvd-2022-22468</content>
    <link href="https://cve.radiocsirt.org/vuln/cnvd-2022-22468"/>
  </entry>
  <entry>
    <id>https://cve.radiocsirt.org/vuln/euvd-2026-28628</id>
    <title>EUVD-2026-28628</title>
    <updated>2026-10-04T16:07:15.753527+00:00</updated>
    <content>EUVD-2026-28628</content>
    <link href="https://cve.radiocsirt.org/vuln/euvd-2026-28628"/>
  </entry>
  <entry>
    <id>https://cve.radiocsirt.org/vuln/fkie_cve-2021-33882</id>
    <title>fkie_cve-2021-33882</title>
    <updated>2026-10-04T16:07:15.753541+00:00</updated>
    <content type="xhtml">
      <div xmlns="http://www.w3.org/1999/xhtml">
        <p>A Missing Authentication for Critical Function vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to reconfigure the device from an unknown source because of lack of authentication on proprietary networking commands.</p>
      </div>
    </content>
    <link href="https://cve.radiocsirt.org/vuln/fkie_cve-2021-33882"/>
  </entry>
  <entry>
    <id>https://cve.radiocsirt.org/vuln/ghsa-7m39-hfcx-7qr8</id>
    <title>GHSA-7m39-hfcx-7qr8</title>
    <updated>2026-10-04T16:07:15.753568+00:00</updated>
    <content type="xhtml">
      <div xmlns="http://www.w3.org/1999/xhtml">
        <p>A Missing Authentication for Critical Function vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to reconfigure the device from an unknown source because of lack of authentication on proprietary networking commands.</p>
      </div>
    </content>
    <link href="https://cve.radiocsirt.org/vuln/ghsa-7m39-hfcx-7qr8"/>
  </entry>
  <entry>
    <id>https://cve.radiocsirt.org/vuln/gsd-2021-33882</id>
    <title>gsd-2021-33882</title>
    <updated>2026-10-04T16:07:15.753583+00:00</updated>
    <content>gsd-2021-33882</content>
    <link href="https://cve.radiocsirt.org/vuln/gsd-2021-33882"/>
  </entry>
  <entry>
    <id>https://cve.radiocsirt.org/vuln/icsma-21-294-01</id>
    <title>ICSMA-21-294-01 — B. Braun Infusomat Space Large Volume Pump</title>
    <updated>2026-10-04T16:07:15.753593+00:00</updated>
    <content type="xhtml">
      <div xmlns="http://www.w3.org/1999/xhtml">
        <p>An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device. CVE-2021-33886 has been assigned to this vulnerability. A CVSS v3 base score of 6.8 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:N). An insufficient verification of data authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data to be used in place of the correct data. This results in full system command access and execution because of the lack of cryptographic signatures on critical data sets. CVE-2021-33885 has been assigned to this vulnerability. A CVSS v3 base score of 9.0 has been calculated; the CVSS vector string is (AV:N/AC:H/PR:N/UI:N/S:C/C:H/I:H/A:H). A missing authentication for critical function vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to reconfigure the device from an unknown source because of the lack of authentication on proprietary networking commands. CVE-2021-33882 has been assigned to this vulnerability. A CVSS v3 base score of 6.8 has been calculated; the CVSS vector string is (AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:H/A:N). A cleartext transmission of sensitive information vulnerability in B. Braun Spac…</p>
      </div>
    </content>
    <link href="https://cve.radiocsirt.org/vuln/icsma-21-294-01"/>
  </entry>
</feed>