<?xml version='1.0' encoding='UTF-8'?>
<?xml-stylesheet href="/static/style.xsl" type="text/xsl"?>
<feed xmlns="http://www.w3.org/2005/Atom" xml:lang="en">
  <id>https://cve.radiocsirt.org/rss/recent/all/10</id>
  <title>Most recent entries from all</title>
  <updated>2026-10-04T01:33:44.178146+00:00</updated>
  <author>
    <name>Vulnerability-Lookup</name>
    <email>csirt@opendfir.org</email>
  </author>
  <link href="https://cve.radiocsirt.org" rel="alternate"/>
  <generator uri="https://lkiesow.github.io/python-feedgen" version="1.0.0">python-feedgen</generator>
  <subtitle>Contains only the most 10 recent entries.</subtitle>
  <entry>
    <id>https://cve.radiocsirt.org/vuln/cnvd-2018-25434</id>
    <title>cnvd-2018-25434</title>
    <updated>2026-10-04T01:33:44.181869+00:00</updated>
    <content>cnvd-2018-25434</content>
    <link href="https://cve.radiocsirt.org/vuln/cnvd-2018-25434"/>
  </entry>
  <entry>
    <id>https://cve.radiocsirt.org/vuln/euvd-2026-69278</id>
    <title>EUVD-2026-69278</title>
    <updated>2026-10-04T01:33:44.181901+00:00</updated>
    <content>EUVD-2026-69278</content>
    <link href="https://cve.radiocsirt.org/vuln/euvd-2026-69278"/>
  </entry>
  <entry>
    <id>https://cve.radiocsirt.org/vuln/fkie_cve-2018-18563</id>
    <title>fkie_cve-2018-18563</title>
    <updated>2026-10-04T01:33:44.181915+00:00</updated>
    <content type="xhtml">
      <div xmlns="http://www.w3.org/1999/xhtml">
        <p>An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial Number below KQ0400000 or KS0400000) and cobas h 232 before 04.00.04 (Serial Number above KQ0400000 or KS0400000). Improper access control to a service command allows attackers in the adjacent network to execute arbitrary code on the system through a crafted Poct1-A message.</p>
      </div>
    </content>
    <link href="https://cve.radiocsirt.org/vuln/fkie_cve-2018-18563"/>
  </entry>
  <entry>
    <id>https://cve.radiocsirt.org/vuln/ghsa-fvc7-cwmq-hcv2</id>
    <title>GHSA-fvc7-cwmq-hcv2</title>
    <updated>2026-10-04T01:33:44.181943+00:00</updated>
    <content type="xhtml">
      <div xmlns="http://www.w3.org/1999/xhtml">
        <p>An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial Number below KQ0400000 or KS0400000) and cobas h 232 before 04.00.04 (Serial Number above KQ0400000 or KS0400000). Improper access control to a service command allows attackers in the adjacent network to execute arbitrary code on the system through a crafted Poct1-A message.</p>
      </div>
    </content>
    <link href="https://cve.radiocsirt.org/vuln/ghsa-fvc7-cwmq-hcv2"/>
  </entry>
  <entry>
    <id>https://cve.radiocsirt.org/vuln/gsd-2018-18563</id>
    <title>gsd-2018-18563</title>
    <updated>2026-10-04T01:33:44.181961+00:00</updated>
    <content>gsd-2018-18563</content>
    <link href="https://cve.radiocsirt.org/vuln/gsd-2018-18563"/>
  </entry>
  <entry>
    <id>https://cve.radiocsirt.org/vuln/icsma-18-310-01</id>
    <title>ICSMA-18-310-01 — Roche Diagnostics Point of Care Handheld Medical Devices (Update A)</title>
    <updated>2026-10-04T01:33:44.181972+00:00</updated>
    <content type="xhtml">
      <div xmlns="http://www.w3.org/1999/xhtml">
        <p>Weak access credentials may enable attackers in the adjacent network to gain unauthorized service access via a service interface. CVE-2018-18561 has been assigned to this vulnerability. A CVSS v3 base score of 6.5 has been calculated; the CVSS vector string is (AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N).. Affected products:CVE-2018-18561 has been assigned to this vulnerability. A CVSS v3 base score of 6.5 has been calculated; the CVSS vector string is (AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N). Insecure permissions in a service interface may allow authenticated attackers in the adjacent network to execute arbitrary commands on the operating systems. CVE-2018-18562 has been assigned to this vulnerability. A CVSS v3 base score of 8.0 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:N/UI:N/S:C/C:N/I:H/A:H).. Affected products:CVE-2018-18562 has been assigned to this vulnerability. A CVSS v3 base score of 8.0 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:N/UI:N/S:C/C:N/I:H/A:H). A vulnerability in the software update mechanism allows an attacker in adjacent network to overwrite arbitrary files on the system through a crafted update package. CVE-2018-18563 has been assigned to this vulnerability. A CVSS v3 base score of 8.0 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:N/UI:N/S:C/C:N/I:H/A:H).. Affected products: Affected products:. Accu-Chek Inform II Instrument - all versions before 03.06.00 (serial number below 14000) / 04.03.00 (serial Number ab…</p>
      </div>
    </content>
    <link href="https://cve.radiocsirt.org/vuln/icsma-18-310-01"/>
  </entry>